Author/Authors :
Yokoyama، نويسنده , , M. and Minamino، نويسنده , , A. and Gomi، نويسنده , , S. and Ieki، نويسنده , , K. and Nagai، نويسنده , , N. and Nakaya، نويسنده , , T. and Nitta، نويسنده , , K. and Orme، نويسنده , , D. and Otani، نويسنده , , M. and Murakami، نويسنده , , T. and Nakadaira، نويسنده , , T. and Tanaka، نويسنده , , M.، نويسنده ,
Abstract :
We have developed a Multi-Pixel Photon Counter (MPPC) for the neutrino detectors of T2K experiment. About 64,000 MPPCs have been produced and tested in about a year. In order to characterize a large number of MPPCs, we have developed a system that simultaneously measures 64 MPPCs with various bias voltage and temperature. The performance of MPPCs are found to satisfy the requirement of T2K experiment. In this paper, we present the performance of 17,686 MPPCs measured at Kyoto University.
Keywords :
quality assurance , Multi-pixel photon counter , Geiger-mode APD , photodetector , characterization