Author/Authors :
Ronzhin، نويسنده , , A. and Albrow، نويسنده , , M.G. and Demarteau، نويسنده , , M. and Los، نويسنده , , S. and Malik، نويسنده , , S. and Pronko، نويسنده , , A. and Ramberg، نويسنده , , E. and Zatserklyaniy، نويسنده , , A.، نويسنده ,
Abstract :
We describe here the development of a time of flight (TOF) system with 10–20 ps resolution for particle identification in a beam line. The detector resolution also was measured with the start and stop counters close together in the 120 GeV proton beam of the Fermilab Test Beam Facility. We tested both microchannel plate photomultipliers (MCP PMT) and silicon photomultipliers (SiPMs), in both cases using Cherenkov light produced in fused silica (quartz) radiators.