Title of article :
Timing performance comparison of digital methods in positron emission tomography
Author/Authors :
Aykac، نويسنده , , Mehmet and Hong، نويسنده , , Inki and Cho، نويسنده , , Sanghee، نويسنده ,
Abstract :
Accurate timing information is essential in positron emission tomography (PET). Recent improvements in high speed electronics made digital methods more attractive to find alternative solutions to create a time mark for an event. Two new digital methods (mean PMT pulse model, MPPM, and median filtered zero crossing method, MFZCM) were introduced in this work and compared to traditional methods such as digital leading edge (LE) and digital constant fraction discrimination (CFD). In addition, the performances of all four digital methods were compared to analog based LE and CFD. The time resolution values for MPPM and MFZCM were measured below 300 ps at 1.6 GS/s and above that was similar to the analog based coincidence timing results. In addition, the two digital methods were insensitive to the changes in threshold setting that might give some improvement in system dead time.
Keywords :
Time resolution , Constant fraction discriminator , Positron emission tomography , Leading edge , Digital timing methods
Journal title :
Astroparticle Physics