Calibration of an electron volt neutron spectrometer
Author/Authors
Mayers، نويسنده , , J. and Adams، نويسنده , , M.A.، نويسنده ,
Pages
10
From page
47
To page
56
Abstract
The procedure for calibrating the VESUVIO eV neutron spectrometer at the ISIS neutron source is described. VESUVIO is used primarily to measure the momentum distribution n(p) of atoms, by inelastic scattering of very high energy (5–150 eV) neutrons. The results of the calibrations show that measurements of n(p) in atoms with masses lower than 16 amu can be measured with a resolution width ∼25% of the intrinsic peak widths in the current instrument configuration. Some suggestions as to how the instrument resolution could be significantly improved are made.