Title of article
ADC common noise correction and zero suppression in the PIBETA detector
Author/Authors
E. Frlez، نويسنده , , E. and Po?ani?، نويسنده , , D. and Ritt، نويسنده , , S.، نويسنده ,
Pages
9
From page
341
To page
349
Abstract
We describe a simple procedure for reducing Analog-to-Digital Converter (ADC) common noise in modular detectors that does not require additional hardware. A method using detector noise groups should work well for modular particle detectors such as segmented electromagnetic calorimeters, plastic scintillator hodoscopes, cathode strip wire chambers, segmented active targets, and the like. We demonstrate a “second pedestal noise correction” method by comparing representative ADC pedestal spectra for various elements of the PIBETA detector before and after the applied correction.
Keywords
Correlated common noise , Data Acquisition (DAQ) zero suppression , ADC pedestals
Journal title
Astroparticle Physics
Record number
2015754
Link To Document