Title of article :
Biparametric correction methods using two shapers for In/CdTe/Pt radiation detector
Author/Authors :
Seino، نويسنده , , Tomoyuki and Ishitsu، نويسنده , , Takafumi and Ueno، نويسنده , , Yuichiro and Kobashi، نويسنده , , Keiji، نويسنده ,
Pages :
5
From page :
170
To page :
174
Abstract :
Biparametric correction methods have reportedly been used to improve the energy resolutions of CdTe or CdZnTe radiation detectors. There are two methods for the correction. One is a method that uses the rise time and pulse height, and the other is a method that uses two pulse heights acquired from a fast and a slow shaper. The latter one for a 2.3-mm-thick In/CdTe/Pt radiation detector was investigated for this paper. The polarization effect resulting in short-term instability in the energy spectrum should be resolved before an In/CdTe/Pt detector is used. A pulsed bias voltage shutdown technique was used with the two shaper biparametric correction method to overcome both the polarization effect and the incomplete carrier collection. An energy resolution of 4.6% for a 122 keV peak was observed for a 2.3-mm-thick In/CdTe/Pt detector at 35 °C for 2 h after applying a bias voltage of −800 V. Furthermore, the energy resolutions are improved for all the photo peaks when a constant correction factor is applied to the five photo peaks in the energy range 59.5–1333 keV.
Keywords :
CdTe , radiation detector , correction method , Energy resolution
Journal title :
Astroparticle Physics