Title of article
Continued development of thallium bromide and related compounds for gamma-ray spectrometers
Author/Authors
Kim، نويسنده , , H. and Churilov، نويسنده , , A. and Ciampi، نويسنده , , G. and Cirignano، نويسنده , , L. and Higgins، نويسنده , , W. and Kim، نويسنده , , S. and O’Dougherty، نويسنده , , P. and Olschner، نويسنده , , F. and Shah، نويسنده , , Kanai، نويسنده ,
Pages
5
From page
192
To page
196
Abstract
Thallium bromide (TlBr) and related ternary compounds, TlBrI and TlBrCl, have been under development for room temperature gamma-ray spectroscopy due to high density, high Z and wide bandgap of the material. Low melting point and cubic crystal structure of selected compositions of these compounds facilitate crystal growth by melt techniques. Recent advances in material purification, crystal growth, and device processing have led to mobility–lifetime products of electrons in the mid 10−3 cm2/V range enabling working detectors of greater than 15 mm thickness to be fabricated.
s paper we report on our recent progress on TlBr detector development and first results from TlBrxCl1−x devices. Pulse height spectra will be presented from TlBr arrays as thick as 18 mm. Depth corrected spectra will also be presented. For a 5 mm thick TlBr array, energy resolution of less than 1% (FWHM at 662 keV) was obtained after depth correction.
Keywords
Gamma-ray spectroscopy , Semiconductor , TlBr
Journal title
Astroparticle Physics
Record number
2015817