Author/Authors :
von Schmid، نويسنده , , M. and Egelhof، نويسنده , , P. and Eremin، نويسنده , , V. and Gernhنuser، نويسنده , , R. and Krِll، نويسنده , , T. and Mutterer، نويسنده , , M. and Pietralla، نويسنده , , N. and Streicher، نويسنده , , B. and Weber، نويسنده , , M.، نويسنده ,
Abstract :
The method of pulse-shape analysis (PSA) for particle identification (PID) was applied to a double-sided silicon strip detector (DSSD) with a strip pitch of 300 μ m . We present the results of test measurements with particles from the reactions of a 70 MeV 12C beam impinging on a mylar target. Good separation between protons and alpha particles down to 3 MeV has been obtained when excluding the interstrip events of the DSSD from the analysis.
Keywords :
DSSSD) , 12C-ions 70 , MEV , Targets: C , Double-sided silicon strip detector (DSSD , Particle identification (PID) , Pulse-shape analysis (PSA) , Mylar