Title of article :
Dedicated scanner for laboratory investigations on cone-beam CT/SPECT imaging of the breast
Author/Authors :
Mettivier، نويسنده , , Giovanni and Russo، نويسنده , , Paolo and Cesarelli، نويسنده , , Mario and Ospizio، نويسنده , , Roberto and Passeggio، نويسنده , , Giuseppe and Roscilli، نويسنده , , Lorenzo and Pontoriere، نويسنده , , Giuseppe and Rocco، نويسنده , , Raffaele، نويسنده ,
Pages :
7
From page :
350
To page :
356
Abstract :
We describe the design, realization and basic tests of a prototype Cone-Beam Breast Computed Tomography (CBBCT) scanner, combined with a SPECT head consisting of a compact pinhole gamma camera based on a photon counting CdTe hybrid pixel detector. The instrument features a 40 μm focal spot X-ray tube, a 50 μm pitch flat panel detector and a 1-mm-thick, 55 μm pitch CdTe pixel detector. Preliminary imaging tests of the separate CT and gamma-ray units are presented showing a resolution in CT of 3.2 mm−1 at a radial distance of 50 mm from the rotation axis and that the 5 and 8 mm hot masses (99mTc labeled with a 15:1 activity ratio with respect to the background) can be detected in planar gamma-ray imaging with a contrast-to-noise ratio of about 4.
Keywords :
Breast computed tomography—SPECT/CT multimodality imaging
Journal title :
Astroparticle Physics
Record number :
2015891
Link To Document :
بازگشت