Author/Authors :
Cattaneo، نويسنده , , P.W. and Argan، نويسنده , , A. and Boffelli، نويسنده , , F. and Bulgarelli، نويسنده , , A. and Buonomo، نويسنده , , B. and Chen، نويسنده , , A.W. and D’Ammando، نويسنده , , F. and Froysland، نويسنده , , T. and Fuschino، نويسنده , , F. San-Galli، نويسنده , , M. and Gianotti، نويسنده , , F. and Giuliani، نويسنده , , A. and Longo، نويسنده , , F. and Marisaldi، نويسنده , , M. and Mazzitelli، نويسنده , , G. and Pel، نويسنده ,
Abstract :
The AGILE scientific instrument has been calibrated with a tagged γ -ray beam at the Beam Test Facility (BTF) of the INFN Laboratori Nazionali di Frascati (LNF). The goal of the calibration was the measure of the Point Spread Function (PSF) as a function of the photon energy and incident angle and the validation of the Monte Carlo (MC) simulation of the silicon tracker operation. The calibration setup is described and some preliminary results are presented.
Keywords :
Telescopes , instrumentation: detectors , gamma rays: observations , Artificial satellites