The stabilization of a positron lifetime spectrometer with a high-accuracy time reference
Author/Authors :
Nissilن، نويسنده , , J. and Karppinen، نويسنده , , M. and Rytsِlن، نويسنده , , K. and Oila، نويسنده , , J. and Saarinen، نويسنده , , K. and Hautojنrvi، نويسنده , , P.، نويسنده ,
Pages :
11
From page :
527
To page :
537
Abstract :
A high-accuracy time reference for a fast coincidence apparatus using scintillation detectors is constructed. The reference peak is created by feeding fast light pulses from a light-emitting diode along two optical fibers of different lengths onto the photomultipliers. The peak can serve as a basis of digital hardware or software stabilization. A particular advantage of the system is that the accuracy of drift detection can be enhanced notably by using a high frequency in the reference peak. In a one-week-long test measurement with a positron lifetime spectrometer, drifts of 12 ps were reduced below a level of 0.5 ps with hardware stabilization.
Keywords :
Fast coincidence apparatus , Digital stabilization , Positron lifetime spectrometer