Title of article :
A complete dosimetric characterization of two 90Sr–90Y dermatologic applicators
Author/Authors :
Coelho، نويسنده , , T.S. and Fernandes، نويسنده , , M.A.R. and Yoriyaz، نويسنده , , H. and Antonio، نويسنده , , P.L.، نويسنده ,
Pages :
5
From page :
147
To page :
151
Abstract :
A complete dosimetric characterization of two Amershan 90Sr–90Y dermatologic applicators is described in this present work. The dosimetric parameters analyzed are: percentage depth dose curve, radial dose distribution, non-uniformity and asymmetry. Both applicators are planar–circular having 22.57 and 9.0 mm diameters. In the range where the percentage depth dose goes from 100% down to 20%, the measured percentage depth dose and that obtained by the Monte Carlo simulation have shown maximum discrepancy of 5.3% for both applicators. The radial dose distribution has been measured at several depths using a GafChromic® EBT QD+ films and it was also calculated by simulation. The discrepancies found did not exceed 5.9% up to the depth of 1.8 mm, where the percentage depth dose drops to 40% of the maximum. The maximum non-uniformity and asymmetry are 1.7% and 5.3% for the first applicator and 22.7% and 25.9% for the second applicator, respectively. Both applicators meet the specification for the maximum non-uniformity established by the adopted protocol, whose limit is 30%. As for the asymmetry the limit is 20% and the second applicator exceeded it in about 5.9%.
Keywords :
Keloids , Betatherapy , dosimetry , Monte Carlo Method , pterygium