• Title of article

    High energy synchrotron X-ray focusing by fixed exit bender

  • Author/Authors

    Yoneda، نويسنده , , Yasuhiro and Matsumoto، نويسنده , , Norimasa and Furukawa، نويسنده , , Yukito and Ishikawa، نويسنده , , Tetsuya، نويسنده ,

  • Pages
    3
  • From page
    370
  • To page
    372
  • Abstract
    The performance of sagittal focusing for hard X-rays with a cylindrical bent crystal at SPring-8 is described. This bending mechanism is designed for the SPring-8 standard double crystal monochromator of bending-magnet beamlines. Therefore, a fixed height exit for a wide energy range (5 – 150 keV) is required. The results underline that focusing in a wide energy band was achieved by combining an adjustable inclined double crystal monochromator and a fixed height exit bending mechanism. The monochromator was maintained at a constant focus at the sample position by optimizing the bent crystal in the energy range from 60 to 150 keV.
  • Keywords
    Fixed exit bender , High energy X-ray , Cylindrical bent crystal
  • Journal title
    Astroparticle Physics
  • Record number

    2016337