Title of article
Optical attenuation measurements in metal-loaded liquid scintillators with a long-pathlength photometer
Author/Authors
Goett، نويسنده , , Johnny and Napolitano، نويسنده , , James K. Yeh، نويسنده , , Minfang and Hahn، نويسنده , , Richard and Cumming، نويسنده , , J.B. and Hans، نويسنده , , Sunej and Rosero، نويسنده , , Richard and Diwan، نويسنده , , Milind and Sexton، نويسنده , , Ken، نويسنده ,
Pages
6
From page
47
To page
52
Abstract
We describe precise optical attenuation measurements of LAB based liquid scintillators using a custom-built photometer with variable pathlengths up to 2 m. Results are presented for linear alkyl benzene (LAB) and LAB based liquid scintillators at wave lengths of 405, 430, 450, and 470 nm. For these measurements much care was exercised to control the purity of the liquids and additives, as well as the materials of the photometer to control contaminations. These data should be valuable for design of large detectors, such as underground neutrino detectors, using these materials as the media.
Keywords
Organic scintillator , Optical attenuation , Neutrino detection , Linear alkyl benzene
Journal title
Astroparticle Physics
Record number
2016342
Link To Document