• Title of article

    PGM beamline with constant energy resolution mode for U49-2 undulator at BESSY-II

  • Author/Authors

    Sawhney، نويسنده , , K.J.S. and Senf، نويسنده , , F. and Gudat، نويسنده , , W.، نويسنده ,

  • Pages
    4
  • From page
    466
  • To page
    469
  • Abstract
    A PGM beamline has been designed for the U49-2 undulator at BESSY-II in which the collimation by the preoptics is done in the dispersive and horizontal direction. The beamline is characterized by high flux, high brightness, high spectral resolution (E/ΔE>5000) over a wide spectral range of ∼70–1500 eV and a relatively small spot size of 70 μm×20 μm (fwhm). To achieve high spectral flux two interchangeable blazed gratings are used in the monochromator. For this beamline a new drive mode has been derived in which almost constant spectral resolution can be maintained while driving the monochromator on-blaze. This constant resolution is maintained even for gratings with different line densities.
  • Keywords
    Plane grating monochromator , Soft X-ray , Undulator beamline , Constant resolution
  • Journal title
    Astroparticle Physics
  • Record number

    2016379