Author/Authors :
Yamazaki، نويسنده , , H. and Yabashi، نويسنده , , M. and Tamasaku، نويسنده , , K. and Yoneda، نويسنده , , Y. and Goto، نويسنده , , S. and Mochizuki، نويسنده , , T. and Ishikawa، نويسنده , , T.، نويسنده ,
Abstract :
A systematic alignment method is described for fixed-exit position operation of SPring-8 standard monochromator in rotated-inclined geometry. Results of high-heat-load test showed the water-cooled pin-post crystal combined with the geometry well surpass 400 W/mm2 power density of the X-ray undulator radiation.
Keywords :
High-heat-load , Rotated-inclined geometry , X-ray monochromator , ALIGNMENT , Cooling Performance