Author/Authors :
Nishibori، نويسنده , , E. and Takata، نويسنده , , M. and Kato، نويسنده , , K. and Sakata، نويسنده , , M. and Kubota، نويسنده , , Y. and Aoyagi، نويسنده , , S. and Kuroiwa، نويسنده , , Y. and Yamakata، نويسنده , , M. and Ikeda، نويسنده , , N.، نويسنده ,
Abstract :
The design and performance of a large Debye–Scherrer Camera with imaging plate (IP) as a detector, which was very recently installed at SPring-8, BL02B2, is reported. By taking advantage of high beam quality of SPring-8, the camera enables one a rapid collection of a high counting statistics and high angular resolution powder pattern, which can lead to accurate structure analyses. The camera also provides easy access to structural changes at varied temperatures between 15–1000 K. The camera provides a rapid and accurate powder diffraction system utilizing third generation SR.