Title of article :
Observations on dual-ended readout of 100 mm long LYSO crystals
Author/Authors :
ur-Rehman، نويسنده , , Fazal and McIntosh، نويسنده , , Bryan and Goertzen، نويسنده , , Andrew L.، نويسنده ,
Pages :
5
From page :
275
To page :
279
Abstract :
We are investigating using dual-ended readout of axially oriented long thin scintillator crystals in detectors for a compact geometry, small ring diameter animal PET system. The axial position of interaction is determined from the light sharing between two photodetectors at opposite ends of the crystal. We examine the light output, energy resolution and axial spatial resolution of 1.5–5×2×100 mm3 polished LYSO crystals by irradiating with an electronically collimated beam of 511 keV photons oriented perpendicular to the long axis and read out at either end by position sensitive photomultiplier tubes (PSPMTs). Three reflector materials, namely Teflon, 3 M enhanced specular reflector (ESR) and black paint are examined for the 2×2×100 mm3 crystal size. The light output increases and energy resolution improves with the crystal cross-section. Generally, the spatial resolution worsens with increase in crystal cross-section. For the 2×2×100 mm3 crystal size, the mean energy resolutions of the photopeak over the nine irradiation positions were 14.4±0.4%, 16.0±1.2% and 28.3±2.1% with mean spatial resolutions of 7.0±1.0, 9.4±3.3 and 26.0±5.0 mm using ESR, Teflon and black paint, respectively. ESR reflector gave the best light output, energy and axial spatial resolutions. These characterization results of PSPMT-based dual-ended long LYSO crystals will be useful in the design of detector modules for a highly compact geometry preclinical PET system using this detector technology.
Keywords :
PSPMT , Axial-positioning , 2×2×100  , mm3 LYSO , DOI , Dual-ended readout , microPET
Journal title :
Astroparticle Physics
Record number :
2017479
Link To Document :
بازگشت