• Title of article

    Characterisation of the IPNE Bonner sphere spectrometer by comparison with the PTB system

  • Author/Authors

    D.J and Alevra، نويسنده , , A.V and Plostinaru، نويسنده , , V.D، نويسنده ,

  • Pages
    5
  • From page
    21
  • To page
    25
  • Abstract
    An existing set of Bonner spheres of the Institute for Physics and Nuclear Engineering in Bucharest has been modified following the PTB design for application of an 3He-filled SP9 counter. By simple interpolations of the fluence responses of the PTB spheres, a preliminary fluence response matrix could be established for the IPNE Bonner sphere set. For further adjustments of the preliminary responses and for validation of the final fluence response matrix of the Romanian Bonner sphere spectrometer, calibration measurements with a reference 252Cf neutron source and joint measurements, including the PTB Bonner sphere spectrometer, were performed in a few neutron fields differing substantially in their spectral distributions. It is estimated that the integral neutron fluences and dose-equivalent values can now be determined with the Romanian spectrometer with uncertainties of about ±4% and ±8%, respectively.
  • Keywords
    Response matrix , Neutron spectrometry , Unfolding , Multisphere neutron spectrometer
  • Journal title
    Astroparticle Physics
  • Record number

    2017999