• Title of article

    Neutron shielding verification measurements and simulations for a 235-MeV proton therapy center

  • Author/Authors

    Newhauser، نويسنده , , W.D and Titt، نويسنده , , U and Dexheimer، نويسنده , , D and Yan، نويسنده , , X and Nill، نويسنده , , S، نويسنده ,

  • Pages
    5
  • From page
    80
  • To page
    84
  • Abstract
    The neutron shielding at the Massachusetts General Hospitalʹs 235-MeV proton therapy facility was investigated with measurements, analytical calculations, and realistic three-dimensional Monte Carlo simulations. In 37 of 40 cases studied, the analytical calculations predicted higher neutron dose equivalent rates outside the shielding than the measured, typically by more than a factor of 10, and in some cases more than 100. Monte Carlo predictions of dose equivalent at three locations are, on average, 1.1 times the measured values. Except at one location, all of the analytical model predictions and Monte Carlo simulations overestimate neutron dose equivalent.
  • Keywords
    neutron shielding , Monte Carlo simulations , Proton therapy
  • Journal title
    Astroparticle Physics
  • Record number

    2018015