Title of article
Design of an improved neutron dose equivalent dosimeter
Author/Authors
Brushwood، نويسنده , , J.M and Beeley، نويسنده , , P.A and Spyrou، نويسنده , , N.M، نويسنده ,
Pages
5
From page
304
To page
308
Abstract
This paper describes the design and development of a new active area neutron dosimeter. The design incorporates a traditional central detector with a moderator/filter arrangement and a number of outer PIN type photodiodes sensitised to thermal neutrons by the application of a lithium fluoride converter. The outer thermal detectors allow the determination of the neutron radiation field characteristics. The experimental programme has demonstrated that such an arrangement is capable of discriminating between various neutron fields and the usefulness of MCNP4b as a design tool.
Keywords
Neutron , dose equivalent , dosimeter
Journal title
Astroparticle Physics
Record number
2018071
Link To Document