Author/Authors :
Abdurashitov، نويسنده , , J.N and Gavrin، نويسنده , , V.N and Kalikhov، نويسنده , , A.V and Matushko، نويسنده , , V.L and Shikhin، نويسنده , , A.A and Yants، نويسنده , , V.E and Zaborskaia، نويسنده , , O.S، نويسنده ,
Abstract :
The principle of operation, design, registration system and main characteristics of a fast neutron spectrometer are described. The spectrometer is intended for direct measurements of ultra low fluxes of fast neutrons. It is sensitive to neutron fluxes of 10−7 cm−2 s−1 and lower. The detection efficiency of fast neutrons with simultaneous energy measurement was determined from Monte-Carlo simulation to be equal to 0.11±0.01. The background counting rate in the detector corresponds to a neutron flux of (6.5±2.1)×10−7 cm−2 s−1 in the range 1.0–11.0 MeV. The natural neutron flux from the surrounding mine rock at the depth of 4600 m of water equivalent was measured to be (7.3±2.4)×10−7 cm−2 s−1 in the range 1.0–11.0 MeV. The flux of fast neutrons in the SAGE main room was measured to be <2.3×10−7 cm−2 s−1 in 1.0–11.0 MeV energy range.
Keywords :
Thermalizer , Helium proportional counters , fast neutrons , Calorimeter , response function , Detection efficiency