Title of article :
Measurement of neutron dose equivalent to proton therapy patients outside of the proton radiation field
Author/Authors :
Yan، نويسنده , , X and Titt، نويسنده , , U and Koehler، نويسنده , , A.M and Newhauser، نويسنده , , W.D، نويسنده ,
Pages :
6
From page :
429
To page :
434
Abstract :
Measurements of neutron dose equivalent values and neutron spectral fluences close to but outside of the therapeutic proton radiation field are presented. The neutron spectral fluences were determined at five locations with Bonner sphere measurements and established by unfolding techniques. More than 50 additional neutron dose equivalent values were measured with LiI and BF3 thermal neutron detectors surrounded by a 25 cm polyethylene moderating sphere. For a large-field treatment, typical values of neutron dose equivalent per therapeutic proton absorbed dose, H/D, at 50 cm distance from isocenter, range from 1 mSv/Gy (at 0° with respect to the proton beam axis) to 5 mSv/Gy (at 90°). Experiments reveal that H/D varies significantly with the treatment technique, e.g., patient orientation, proton beam energy, and range-modulation. The relative uncertainty in H/D values is approximately 40% (one standard deviation).
Keywords :
Proton therapy , Patient neutron exposures