Title of article :
PRIMA: An apparatus for medical application
Author/Authors :
Sipala، نويسنده , , V. and Brianzi، نويسنده , , M. and Bruzzi، نويسنده , , M. and Bucciolini، نويسنده , , M. and Cirrone، نويسنده , , G.A.P. and Civinini، نويسنده , , C. and Cuttone، نويسنده , , G. and Lo Presti، نويسنده , , D. and Pallotta، نويسنده , , S. and Randazzo، نويسنده , , Dimitris N. and Romano، نويسنده , , F. and Stancampiano، نويسنده , , C. and Scaringella، نويسنده , , M. and Talamonti، نويسنده , , C. and Tesi، نويسنده , , M.، نويسنده ,
Pages :
5
From page :
73
To page :
77
Abstract :
In this paper a proton Computed Radiography (pCR) apparatus for medical applications, realized by PRIMA (PRoton IMAging) Italian Collaboration, is described. The system is oriented to acquire tomography images and meets clinical demands for the use of protons in radiotherapy treatments. The approach proposed here is based on ‘single proton tracking’ method with Most Likely Path (MLP) reconstruction of the single particle. A pCR prototype, with a field of view of about 5×5 cm2 and an acquisition time of the order of 10 s (10 kHz, 105 events), has been developed and tested with a 62 MeV proton beam at the INFN-Laboratori Nazionali del Sud (LNS). The apparatus architecture will be described and first proton radiographies will be shown.
Keywords :
Proton imaging , Silicon tracker , Calorimeter , Proton therapy
Journal title :
Astroparticle Physics
Record number :
2018402
Link To Document :