• Title of article

    High rate X-ray imaging using multi-GEM detectors with a novel readout design

  • Author/Authors

    Bachmann، نويسنده , , S and Kappler، نويسنده , , S and Ketzer، نويسنده , , B and Müller، نويسنده , , Th and Ropelewski، نويسنده , , L and Sauli، نويسنده , , F and Schulte، نويسنده , , E، نويسنده ,

  • Pages
    5
  • From page
    104
  • To page
    108
  • Abstract
    Modern micropattern gas counters are able to detect soft X-rays with high spatial resolution. Based on the principle of gas amplification, this detector type has a wide dynamic range and provides energy selection during image reconstruction. One of the most recent developments is the Multi-GEM, an efficient, high-rate capable micropattern gas detector. Owing to the separation of gas amplification and readout stage, this detector allows high flexibility in the geometry of the readout structure. two-dimensional projective readout, GEM detectors have shown good capability in providing X-ray absorption radiographies of small mammals. We present a detector with a novel design where the readout board consists of micropads connected to three layers of strips in such a way that two-dimensional spatial reconstruction and an unambiguous resolution of events with multiple hits is possible.
  • Keywords
    x-ray imaging , Gas electron multiplier , GEM , Micropad readout , High rate radiation detector , Unambiguous resolution , Multiple hits
  • Journal title
    Astroparticle Physics
  • Record number

    2018407