Title of article
The response of a liquid scintillator detector to 21–100 MeV neutrons
Author/Authors
Thun، نويسنده , , J. and Blomgren، نويسنده , , J. and Elmgren، نويسنده , , K. and Kنllne، نويسنده , , J. B. Olsson، نويسنده , , N. and Lecolley، نويسنده , , J.F. and Lefebvres، نويسنده , , F. and Varignon، نويسنده , , R. C. Le Borne، نويسنده , , F. and Ledoux، نويسنده , , X. and Patin، نويسنده , , Y. and Jonsson، نويسنده , , O. and Renberg، نويسنده , , P.-U.، نويسنده ,
Pages
18
From page
559
To page
576
Abstract
A large volume (4 l) liquid NE213 scintillator detector for neutrons in the energy range up to 800 MeV has been calibrated using a high accuracy tagging technique at En=21,34,50,60 and 100 MeV. The experiment is described together with a discussion of the accuracy in the absolute efficiency calibration. Results are presented on total and differential detection efficiencies along with information on the underlying nuclear reactions causing the detector response, including both charged particles and γ-rays. The present results complement previous measurements for the same detector, so that an efficiency calibration now exists over an energy range of En=2–800 MeV. The empirical results are compared with a calculated energy dependence of the detector response to neutrons.
Keywords
Liquid scintillator , neutron detection , tagging , Neutron detector efficiency , Spatial efficiency , Neutron beam
Journal title
Astroparticle Physics