Title of article
Detection of krypton in xenon for dark matter applications
Author/Authors
Dobi، نويسنده , , A. and Davis، نويسنده , , C. and Hall، نويسنده , , C. and Langford، نويسنده , , T. and Slutsky، نويسنده , , S. and Yen، نويسنده , , Y.-R.، نويسنده ,
Pages
6
From page
1
To page
6
Abstract
We extend our technique for observing very small concentrations of impurities in xenon gas to the problem of krypton detection. We use a conventional mass spectrometer to identify the krypton content of the xenon, but we improve the sensitivity of the device by more than five orders of magnitude with a liquid nitrogen cold trap. We find that the absolute krypton concentration in the xenon can be inferred from the mass spectrometry measurements, and we identify krypton signals at concentrations as low as 0.5×10−12 mol/mol (Kr/Xe). This technique simplifies the monitoring of krypton backgrounds for WIMP dark matter searches in liquid xenon.
Keywords
Cold trap , krypton , mass spectrometry , Xenon , Dark matter
Journal title
Astroparticle Physics
Record number
2018704
Link To Document