High-rate counters versus integrating devices in synchrotron experiments
Author/Authors
Bordessoule، نويسنده , , Michel and Lemonnier، نويسنده , , Marc، نويسنده ,
Pages
9
From page
592
To page
600
Abstract
The purpose of this paper is to illustrate the finite precision of counting measurements, due to the Poisson distribution of the photons from a pulsed source. This limitation of precision is important when the count rate is of the same order of magnitude as the bunch rate. We compare analytic results with measurements from a counting experiment using an avalanche photodiode and an integrating detector. Simple formulas are given stating both the dynamic and precision of counting with a periodic source.
Keywords
Dead time correction , Poisson Distribution , X-ray counter , Synchrotron radiation , Avalanche diode , Pulse pile up