Title of article :
Refinement of position resolution in two-dimensional X-ray detector based on gaseous detector
Author/Authors :
Hattori، نويسنده , , K. and Parker، نويسنده , , J.D. and Ida، نويسنده , , C. and Ito، نويسنده , , K. and Kubo، نويسنده , , H. and Miuchi، نويسنده , , K. and Tanimori، نويسنده , , T. and Takata، نويسنده , , M.، نويسنده ,
Pages :
7
From page :
1
To page :
7
Abstract :
We have successfully refined the position resolution of a two-dimensional X-ray photon-counting detector based on the micro-pixel gas chamber ( μ - PIC ) by measuring the charge distribution of an X-ray interaction without the use of analog-to-digital converters (ADCs). By updating the logic of the Field Programmable Gate Arrays (FPGAs) included in the data acquisition system, we were able to acquire the pulse widths, or time-above-threshold of the μ - PIC signals, by measuring both the leading and trailing edges of the digital signals. By using the measured widths to estimate the peak of the charge distribution, the position resolution of our detector was improved to σ = 93.3 ± 2.8 μ m from a value of σ = 229.5 ± 6.8 μ m found using the FPGA logic of the previous system. This represents an improvement of nearly 60%.
Keywords :
Gaseous detector , Micro-pattern detector , X-ray detector