Title of article :
Dose responses in a normoxic polymethacrylic acid gel dosimeter using optimal CT scanning parameters
Author/Authors :
Cho، نويسنده , , Maylor K.H. and Cho، نويسنده , , S.J. and Lee، نويسنده , , S. and Lee، نويسنده , , S.H. and Min، نويسنده , , C.K. and Kim، نويسنده , , Y.H. and Moon، نويسنده , , S.K and Kim، نويسنده , , E.S. and Chang، نويسنده , , A.R. and Kwon، نويسنده , , S.I.، نويسنده ,
Pages :
6
From page :
112
To page :
117
Abstract :
The dosimetric characteristics of normoxic polymethacrylic acid gels are investigated using optimal CT scanning parameters and the possibility of their clinical application is also considered. The effects of CT scanning parameters (tube voltage, tube current, scan time, slick thickness, field of view, and reconstruction algorithm) are experimentally investigated to determine the optimal parameters for minimizing the amount of noise in images obtained using normoxic polymethacrylic acid gel. In addition, the dose sensitivity, dose response, accuracy, and reproducibility of the normoxic polymethacrylic acid gel are evaluated. ges are obtained using a head phantom that is fabricated for clinical applications. In addition, IMRT treatment planning is performed using a Tomotherapy radiation treatment planning system. A program for analyzing the results is produced using Visual C. A comparison between the treatment planning and the CT images of irradiated gels is performed. The dose sensitivity is found to be 2.41±0.04 HGy−1. The accuracies of dose evaluation at doses of 2 Gy and 4 Gy are 3.0% and 2.6%, respectively, and their reproducibilities are 2.0% and 2.1%, respectively. In the comparison of gel and Tomotherpay planning, the pass rate of the γ-index, based on the reference values of a dose error of 3% and a DTA of 3 mm, is 93.7%.
Keywords :
X-ray CT , CT scanning parameters , Normoxic polymethacrylic acid gel
Journal title :
Astroparticle Physics
Record number :
2019276
Link To Document :
بازگشت