Author/Authors :
?irkovi?، نويسنده , , Sa?a and Ili?، نويسنده , , Andjelija ?. and Dobrosavljevi?، نويسنده , , Aleksandar and Balvanovi?، نويسنده , , Roman and Risti?-Djurovi?، نويسنده , , Jasna L.، نويسنده ,
Abstract :
Systematic errors caused by the deficiencies of the measurement equipment are occurring relatively often in the engineering practice. The magnetic field measurement system of the VINCY Cyclotron has been designed well; however, due to the practical limitations in the machining process fabrication errors are imminent. We present our experiences with the produced measurement system and the techniques used for the detection and correction of errors. A change in the measuring protocol is suggested in order to overcome otherwise unavoidable errors caused by measuring system machining imperfections.
Keywords :
Cyclotrons , Ion beams , Magnetic field measurement , Systematic error , Data analysis