• Title of article

    Metal micro-detector TimePix imaging synchrotron radiation beams at the ESRF Bio-Medical Beamline ID17

  • Author/Authors

    Pugatch، نويسنده , , V. and Campbell، نويسنده , , M. and Chaus، نويسنده , , A. and Kovalchuk، نويسنده , , O. and Llopart، نويسنده , , X. and Okhrimenko، نويسنده , , O. and Pospisil، نويسنده , , S. and Prezado، نويسنده , , Y. and Renier، نويسنده , , M. and Tlustos، نويسنده , , L.، نويسنده ,

  • Pages
    4
  • From page
    8
  • To page
    11
  • Abstract
    Characterization studies of the metal micro-detector TimePix measuring dose distribution at the Minibeam Radiation Therapy setup (Bio-Medical Beamline ID17, ESRF) were performed. The results obtained for high intensity synchrotron radiation minibeams illustrate an excellent performance of the TimePix providing in real time 2D image of the dose distribution over many beams in a 14×14 mm2 area. Peak-to-Valley–Dose–Ratios measured by TimePix and gafhromic films agree well.
  • Keywords
    Pixel detector , Metal micro-detector , radiation therapy , Real time 2D-dose distribution monitoring , Timepix , Peak-to-Valley–Dose–Ratio
  • Journal title
    Astroparticle Physics
  • Record number

    2019466