Title of article :
Quality assessment of neutron delivery system for small-angle neutron scattering diffractometers of the Jülich Centre for Neutron Science at the FRM II
Author/Authors :
Radulescu، نويسنده , , Aurel and Pipich، نويسنده , , Vitaliy and Ioffe، نويسنده , , Alexander، نويسنده ,
Abstract :
Following the shutdown of FRJ-2 research reactor in Jülich, the pinhole small-angle neutron diffractometers KWS-1 and KWS-2 have been moved to the research reactor FRM II in Garching. The installation of these 40 m long instruments required the design and setup of new neutron guides with geometrical and optical features imposed by the instrumentsʹ positioning in the neutron guide hall, such as, the predetermined length and beam height as well as the foreseen improvement of the instrument performance. We report here about the quality assessment of the newly constructed neutron guides with respect to the optical, geometrical and alignment characteristics and the positioning of the velocity selector integrated in the neutron guide system by comparing the features of the measured neutron beams (in terms of neutron flux, intensity distribution and beam profile) with the results of the simulations of optimal neutron guide systems.
Keywords :
Neutron guide , Monte-Carlo simulations , Small-angle neutron scattering
Journal title :
Astroparticle Physics