Author/Authors :
Stevanato، نويسنده , , L. and Cester، نويسنده , , D. and Nebbia، نويسنده , , G. and Viesti، نويسنده , , G.، نويسنده ,
Abstract :
Using a fast digitizer, the neutron–gamma discrimination capability of the new liquid scintillator EJ-309 is compared with that obtained using standard EJ-301. Moreover the capability of both the scintillation detectors to identify a weak neutron source in a high gamma-ray background is demonstrated. The probability of neutron detection is PD=95% at 95% confidence level for a gamma-ray background corresponding to a dose rate of 100 μSv/h.