• Title of article

    Cleaning and recirculation of perfluorohexane (C6F14) in the STAR-RICH detector

  • Author/Authors

    Andres، نويسنده , , Y. and Braem، نويسنده , , A. and Cozza، نويسنده , , D. and Davenport، نويسنده , , M. and De Cataldo، نويسنده , , G. and Dell Olio، نويسنده , , L. and DiBari، نويسنده , , D. and DiMauro، نويسنده , , A. and Dunlop، نويسنده , , J.C. and Finch، نويسنده , , E. and Fraissard، نويسنده , , D. and Franco، نويسنده , , A. and Gans، نويسنده , , J. and Ghidini، نويسنده , , B. and Harris، نويسنده , , J.W. and Horsley، نويسنده , , M. and ، نويسنده ,

  • Pages
    20
  • From page
    590
  • To page
    609
  • Abstract
    A RICH detector with a CsI photo-cathode and liquid perfluorohexane radiator has been installed in the STAR experiment at RHIC. The liquid is continuously cleaned and distributed to a quartz containment vessel within the detector by a closed recirculation system. A VUV spectrometer is connected to the system which monitors the optical transparency of the liquid. This measurement provides one of the pieces of information necessary to model the number of Cherenkov photons which reach the pad plane. A description of the liquid recirculation system and the cleaning procedure for the liquid as well as the spectrometer is presented along with results of their performance.
  • Journal title
    Astroparticle Physics
  • Record number

    2019947