Title of article :
Absolute dose calibration of an X-ray system and dead time investigations of photon-counting techniques
Author/Authors :
Carpentieri، نويسنده , , C and Schwarz، نويسنده , , C and Ludwig، نويسنده , , J and Ashfaq، نويسنده , , A and Fiederle، نويسنده , , M، نويسنده ,
Abstract :
High precision concerning the dose calibration of X-ray sources is required when counting and integrating methods are compared.
se calibration for a dental X-ray tube was executed with special dose calibration equipment (dosimeter) as function of exposure time and rate. Results were compared with a benchmark spectrum and agree within ±1.5%.
ime investigations with the Medipix1 photon-counting chip (PCC) have been performed by rate variations. Two different types of dead time, paralysable and non-paralysable will be discussed. The dead time depends on settings of the front-end electronics and is a function of signal height, which might lead to systematic defects of systems. Dead time losses in excess of 30% have been found for the PCC at 200 kHz absorbed photons per pixel.
Keywords :
Medipix , MEDICAL IMAGING , X-rays , Dead time , Dose calibration , Photon-counting chip
Journal title :
Astroparticle Physics