Uozumi، نويسنده , , S and Asakawa، نويسنده , , T and Done، نويسنده , , J.P and Fujii، نويسنده , , Y and Furukawa، نويسنده , , K and Hara، نويسنده , , K and Ishizaki، نويسنده , , T and Kaga، نويسنده , , S and Kajino، نويسنده , , F and Kamon، نويسنده , , T and Kanaya، نويسنده , , N and Kanzaki، نويسنده , , J and Kawagoe، نويسنده , , K and Kim، نويسنده , , S and Miyajima، نويسنده , , A and Nakagawa، نويسنده , , A and Nozaki، نويسنده , , M and Oishi، نويسنده , , R and Ot، نويسنده ,
Pages :
17
From page :
291
To page :
307
Abstract :
We have measured performance of a lead/plastic scintillator sampling calorimeter in two separate beam tests at low (1–4 GeV) and high (10–200 GeV) energies. The calorimeter is composed of 8-mm-thick lead plates and 2-mm-thick plastic scintillator plates for hardware compensation, where responses to electromagnetic and hadronic showers of the same energy are identical. We find the linearity to be better than 1% in the energy range between 2 and 150 GeV for both pions and electrons. The energy resolutions are obtained to be (46.7±0.6)%/E⊕(0.9±0.9)% for pions, where the energy E is given in GeV. The response ratio of electromagnetic showers to hadronic showers is measured to be 1.04±0.01 at low energies, and 0.99±0.01 at high energies.
Keywords :
Calorimeter , Hadron calorimeter , Tile/fiber , compensation , e/?