• Title of article

    Fast beam cut-off method in RF-knockout extraction for spot-scanning

  • Author/Authors

    Furukawa، نويسنده , , Takuji and Noda، نويسنده , , Koji، نويسنده ,

  • Pages
    9
  • From page
    59
  • To page
    67
  • Abstract
    An irradiation method with magnetic scanning has been developed in order to provide accurate irradiation even for an irregular target shape. The scanning method has strongly required a lower ripple of the beam spill and a faster response to beam-on/off in slow extraction from a synchrotron ring. At HIMAC, RF-knockout extraction has utilized a bunched beam to reduce the beam-spill ripple. Therefore, particles near the resonance can be spilled out from the separatrices by synchrotron oscillation as well as by a transverse RF field. From this point of view, a fast beam cut-off method has been proposed and verified by both simulations and experiments. The maximum delay from the beam cut-off signal to beam-off has been improved to around 60 μs from 700 μs by a usual method. Unwanted dose has been considerably reduced by around a factor of 10 compared with that by the usual method.
  • Keywords
    RF-knockout slow extraction , Heavy-ion therapy , Synchrotron , Spot-scanning , Dose management
  • Journal title
    Astroparticle Physics
  • Record number

    2020281