Title of article :
A purity monitoring system for the H1 liquid argon calorimeter
Author/Authors :
Barrelet، نويسنده , , E. and Andrieu، نويسنده , , B. and Babaev، نويسنده , , A. and Banas، نويسنده , , E. and Bederede، نويسنده , , D. and Biddulph، نويسنده , , P. and Borras، نويسنده , , K. and Brasse، نويسنده , , F. and Brisson، نويسنده , , V. and Burmeister، نويسنده , , P. and Buschhorn، نويسنده , , G. and Canton، نويسنده , , B. and Cornett، نويسنده , , U. and Cozzika، نويسنده , , G. and Cvach، نويسنده , , J. and Cyz، نويسنده , , A. and Dar، نويسنده ,
Pages :
19
From page :
204
To page :
222
Abstract :
The ionization probes used for monitoring the liquid argon purity in the H1 calorimeter are described and results of their operation in tests at CERN and during the period 1992 to the end of 1998 at HERA are given. The high sensitivity of the charge measurements leads to refined charge collection models, and to the observation of a variation of the ionization yield of our electron sources with temperature.
Keywords :
Liquid argon calorimeter , Purity , Monitor , Thermal effect , attachment