Title of article
Balloon flight test of pulse shape discrimination (PSD) electronics and background model performance on the HIREGS payload
Author/Authors
Boggs، نويسنده , , S.E and Jean، نويسنده , , P and Slassi-Sennou، نويسنده , , S and Coburn، نويسنده , , W and Lin، نويسنده , , R.P. and Madden، نويسنده , , N.W. and McBride، نويسنده , , S and Pelling، نويسنده , , R.M and Primbsch، نويسنده , , J.H and von Ballmoos، نويسنده , , P، نويسنده ,
Pages
12
From page
390
To page
401
Abstract
We flew a prototype of the INTEGRAL/SPI PSD electronics with a narrow-bore coaxial Ge γ-ray detector on the January 1998 Antarctic balloon flight of the HIREGS payload. The first goal for this test flight was to demonstrate the performance of the PSD electronics in a radiation background and space environment similar to that in which SPI will operate, which was successfully demonstrated. The second goal was to measure the fraction of the background due to localized β−-decays, and to validate the model used to predict SPIʹs background and the sensitivity improvement due to the PSD background rejection. Our background measurements show good agreement with the models developed for SPI, and strongly support the expected improvement in sensitivity due to PSD on SPI.
Keywords
Detectors , Gamma Ray , Germanium
Journal title
Astroparticle Physics
Record number
2020552
Link To Document