Title of article :
Belle electromagnetic calorimeter
Author/Authors :
Miyabayashi، نويسنده , , Kenkichi، نويسنده ,
Pages :
5
From page :
298
To page :
302
Abstract :
We report the performance of the Belle electromagnetic calorimeter for the first three years operation. Good mass resolutions for π0 and η are obtained to be 4.8 and 12.1 MeV/c2, respectively. The degradation of light output due to the radiation damage is small, about 3% for the radiation dose of 40 rad. These performances promise further study of B physics with neutral particle reconstruction.
Keywords :
Belle detector , Electromagnetic calorimetry , CsI(Tl) crystal
Journal title :
Astroparticle Physics
Record number :
2020732
Link To Document :
بازگشت