Author/Authors :
Bezzubov، نويسنده , , V and Denisov، نويسنده , , S and Erin، نويسنده , , S and Ferapontov، نويسنده , , A and Gilitsky، نويسنده , , Yu and Korablev، نويسنده , , V and Kurchaninov، نويسنده , , L and Lobanov، نويسنده , , M and Rybin، نويسنده , , A and Solin، نويسنده , , A and Suzdalev، نويسنده , , V and Tikhonov، نويسنده , , V، نويسنده ,
Abstract :
The performance of a gas ionization electromagnetic calorimeter with planar electrodes and lead absorbers has been investigated using 10–40 GeV electron beams of the IHEP accelerator. The results of signal and noise spectra measurements vs. electron energy, high voltage and gas pressure are presented. The energy dependence of energy resolution is compared with Monte Carlo simulations.