• Title of article

    The use of Monte Carlo calculations in the determination of a Ge detector efficiency curve

  • Author/Authors

    Helmer، نويسنده , , R.G and Hardy، نويسنده , , J.C and Iacob، نويسنده , , V.E and Sanchez-Vega، نويسنده , , M and Neilson، نويسنده , , R.G and Nelson، نويسنده , , J، نويسنده ,

  • Pages
    22
  • From page
    360
  • To page
    381
  • Abstract
    Precise measured data and an extensive set of Monte Carlo calculations have been combined to optimize the parameters for a 280 cm3 n-type Ge semiconductor photon detector, with the outcome that the Monte Carlo calculations now provide a very precise detector efficiency curve. The detector position and its length were determined from a set of measurements that included an axial scan. Measurements of the relative efficiencies based on radionuclides with accurately known relative photon emission rates were combined with efficiencies calculated by a Monte Carlo photon and electron transport code to adjust the detector diameter, internal deadlayer thickness, and the effective external deadlayer thickness. Our results show that, in a well-studied situation, it is possible to use Monte Carlo calculations to aid in the determination of a Ge detectorʹs efficiency with an accuracy of about 0.2% from 50 to 1400 keV.
  • Keywords
    Ge detector , efficiency , Monte Carlo calculations
  • Journal title
    Astroparticle Physics
  • Record number

    2021219