• Title of article

    Evaluation of flat panel PMT for gamma ray imaging

  • Author/Authors

    Pani، نويسنده , , R. and Cinti، نويسنده , , M.N. and Pellegrini، نويسنده , , R. and Trotta، نويسنده , , C. and Trotta، نويسنده , , G. and Montani، نويسنده , , L. and Ridolfi، نويسنده , , S. and Garibaldi، نويسنده , , F. and Scafè، نويسنده , , R. and Belcari، نويسنده , , N. and Del Guerra، نويسنده , , A.، نويسنده ,

  • Pages
    7
  • From page
    262
  • To page
    268
  • Abstract
    The first position sensitive PMT, Hamamatsu R2486, developed in 1985, represented a strong technological advance for gamma-ray imaging. Hamamatsu H8500 Flat Panel PMT is the last generation position sensitive PMT: extremely compact with 2 in. active area. Its main features are: minimum peripheral dead zone (1 mm) and height of 12 mm. It was designed to be assembled in array to cover large detection area. It can represent a technical revolution for many applications in the field of gamma-ray imaging as for example nuclear medicine. This tube is based on metal channel dynode for charge multiplication and 8×8 anodes for charge collection and position calculation. In this paper we present a preliminary evaluation of the imaging performances addressed to nuclear medicine application. s aim we have taken into account two different electronic readouts: resistive chain with Anger Camera principle and multianode readout. Flat panel PMT was coupled to CsI(Tl) and NaI(Tl) scintillation arrays. The results were also compared with the first generation PSPMT.
  • Keywords
    nuclear medicine imaging , Position sensitive photomultiplier , Scintillation array , Flat panel photomultiplier
  • Journal title
    Astroparticle Physics
  • Record number

    2021684