Title of article
Study and application of low pressure multi-wire proportional chambers
Author/Authors
Ye، نويسنده , , Y.L. and Di، نويسنده , , Z.Y. and Li، نويسنده , , Z.H. and Wang، نويسنده , , Q.J and Zheng، نويسنده , , T. and Chen، نويسنده , , T. and Jiang، نويسنده , , D.X. and Ge، نويسنده , , William Y.C. and Pang، نويسنده , , D.Y. and Li، نويسنده , , X.Q.، نويسنده ,
Pages
7
From page
718
To page
724
Abstract
Low Pressure Multi-Wire Proportional Chambers (LPMWPC) were made in the laboratory and the systematic tests were carried out. The position resolution of the chamber is about 0.5 mm and the time resolution is about 1.5 ns. The detection efficiency is higher than 90% for light nuclei at low and intermediate energies, and for beam intensities lower than 105 pps. This kind of detector is applicable for monitoring the radioactive ion beam which is in general characterized by low intensity and large spot. Application of new read-out techniques may allow to improve the beam intensity capability.
Keywords
LPMWPC , Time resolution , Detection efficiency , Position resolution
Journal title
Astroparticle Physics
Record number
2022017
Link To Document