• Title of article

    Gamma spectroscopy using two Clover detectors in close geometry

  • Author/Authors

    Dababneh، نويسنده , , S and Patronis، نويسنده , , N and Assimakopoulos، نويسنده , , P.A and Gِrres، نويسنده , , J and Heil، نويسنده , , M and Kنppeler، نويسنده , , F and Karamanis، نويسنده , , D and OʹBrien، نويسنده , , S and Reifarth، نويسنده , , R، نويسنده ,

  • Pages
    10
  • From page
    230
  • To page
    239
  • Abstract
    A gamma spectrometer composed of two Clover Ge detectors in close geometry is described. The advantages and drawbacks of the different modes of operation are investigated. The use of offline coincidence analysis for substantial background reduction is presented and an experimental approach for the determination of the summing correction factor is formulated.
  • Keywords
    Direct mode , Background reduction , Clover detector , gamma-ray detection , Summing correction , Addback mode , Coincidence mode