Title of article :
Gamma spectroscopy using two Clover detectors in close geometry
Author/Authors :
Dababneh، نويسنده , , S and Patronis، نويسنده , , N and Assimakopoulos، نويسنده , , P.A and Gِrres، نويسنده , , J and Heil، نويسنده , , M and Kنppeler، نويسنده , , F and Karamanis، نويسنده , , D and OʹBrien، نويسنده , , S and Reifarth، نويسنده , , R، نويسنده ,
Pages :
10
From page :
230
To page :
239
Abstract :
A gamma spectrometer composed of two Clover Ge detectors in close geometry is described. The advantages and drawbacks of the different modes of operation are investigated. The use of offline coincidence analysis for substantial background reduction is presented and an experimental approach for the determination of the summing correction factor is formulated.
Keywords :
Direct mode , Background reduction , Clover detector , gamma-ray detection , Summing correction , Addback mode , Coincidence mode
Journal title :
Astroparticle Physics
Record number :
2022223
Link To Document :
بازگشت