Title of article :
First results from a dispersive EXAFS beamline developed at INDUS-2 synchrotron source at RRCAT, Indore, India
Author/Authors :
Bhattacharyya، نويسنده , , D. and Poswal، نويسنده , , A.K. and Jha، نويسنده , , S.N. and Sangeeta and Sabharwal، نويسنده , , S.C.، نويسنده ,
Pages :
8
From page :
286
To page :
293
Abstract :
A dispersive extended X-ray absorption fine structure (EXAFS) beamline employing an elliptically bent crystal and a position-sensitive CCD detector has been developed at the BL-8 bending magnet port of INDUS-2 synchrotron source at RRCAT, Indore, India. The beamline enables simultaneous measurement of X-ray absorption spectra over the range of 5–20 keV in a short span of time. The beamline performance has been investigated by measuring the absorption spectra of metal foils and oxide powders. The EXAFS measurements on crystalline Pb5Ge3O11 samples were performed for the first time in literature. The average Pb–O bond length obtained from these measurements is found to agree well with the theoretically estimated value.
Keywords :
Dispersive EXAFS , Synchrotron beamline , PBO , Bi2O3 , Pb5Ge3O11
Journal title :
Astroparticle Physics
Record number :
2022642
Link To Document :
بازگشت