Title of article :
A pixel chamber to monitor the beam performances in hadron therapy
Author/Authors :
Bonin، نويسنده , , R. and Boriano، نويسنده , , A. and Bourhaleb، نويسنده , , F. and Cirio، نويسنده , , R. and Donetti، نويسنده , , M. and Garelli، نويسنده , , E. and Giordanengo، نويسنده , , S. and Marchetto، نويسنده , , F. and Peroni، نويسنده , , C. and Sanz Freire، نويسنده , , C.J. and Simonetti، نويسنده , , L.، نويسنده ,
Pages :
13
From page :
674
To page :
686
Abstract :
In this paper we describe the design, construction, and tests of a parallel plate ionization chamber with the anode segmented in (32×32) square pixels. The performance of the read out and data acquisition systems is also discussed. The design of the chamber has been finalized to be used as a beam monitor for therapeutical treatments. Position and flux resolution obtained with a carbon ion beam are presented.
Keywords :
radiotherapy , Beam monitor , ionization chamber , dosimetry , Carbon ion