Title of article :
Characterization of an Al-STJ-based X-ray detector with monochromatized synchrotron radiation
Author/Authors :
Huber، نويسنده , , M. and Bechstein، نويسنده , , S. and Beckhoff، نويسنده , , B. and Feilitzsch، نويسنده , , F.v. and Jochum، نويسنده , , J. and Krumrey، نويسنده , , M. and Rüdig، نويسنده , , A. and Ulm، نويسنده , , G.، نويسنده ,
Pages :
3
From page :
234
To page :
236
Abstract :
The characterization of an Al-STJ-based detector with Pb absorber was performed with monochromatized synchrotron radiation. Detector response was measured in the energy range from 3 to 10 keV. A small non-linearity of the signal pulse height was detected, probably due to the escape of recombination phonons from the detector. The non-linearity can be described by a second order polynomial function. Additionally, detector signals were recorded while an X-ray beam of 50 μm diameter was directed to several locations on and near the absorber. For a well-aligned beam, detector artefacts are of at least two orders of magnitude lower intensity than the absorber events.
Keywords :
X-ray detector , Synchrotron radiation , Superconducting tunnel junction , STJ
Journal title :
Astroparticle Physics
Record number :
2022881
Link To Document :
بازگشت