Title of article :
Analyses on the operating point dependence of the energy resolution with a Ti/Au TES microcalorimeter
Author/Authors :
Morita، نويسنده , , U and Ishisaki، نويسنده , , Y and Koga، نويسنده , , T and Ohashi، نويسنده , , T and Terracol، نويسنده , , S.F and Miyazaki، نويسنده , , T and Friedrich، نويسنده , , S and Mitsuda، نويسنده , , K and Tanaka، نويسنده , , K، نويسنده ,
Pages :
4
From page :
277
To page :
280
Abstract :
We examined the performance of a single pixel Ti/Au transition-edge sensor (TES) calorimeter for incident X-ray energies of Al-Kα, Cr-Kα, and Fe-Kα, as a function of the TES resistance. We find that the energy resolution does not always degrade with increasing energy. The best energy resolution of 5.7±0.9 eV at 6.4 keV is obtained, which is possibly even better than the baseline width of 6.5±0.2 eV. Assuming that the noise level is determined by the noise spectrum NS(f;R∼R+dR(E)) considering the resistance change of dR(E), instead of NS(f;R) at the operating point, these results may be explained by the fact that the noise decreases at the higher TES resistance. The pulse variation appears to have a minimum at a certain resistance of R+dR(E)≃48 mΩ, and the best energy resolution for each line is obtained at such an operating point, respectively. The pulse variation could be enhanced when the fluctuation of the TES sensitivity α is large at R+dR(E).
Keywords :
X-ray detectors , Microcalorimeter , Superconducting transition edge sensor
Journal title :
Astroparticle Physics
Record number :
2022894
Link To Document :
بازگشت